Chapters Transcript 3 Sensica™ Urine Output System Common Procedures If there is visible urine in the tubing above the system, drain it into the bag by lifting the tube while keeping the tube attached to the monitor allow the urine output to be recorded by the sensical urine output system, next take the bag off the ring and drain it's not necessary to remove the tubing from the tube holder. A message will appear that says the bag has been removed after draining the bag, hang it back on the ring, touch the resume button. The system will take a few seconds to stabilize. Then resume monitoring urine output monitoring will automatically resume after calibration with the bag level at 0%. Note the current. Our volume remains unchanged as this captures the patient's cumulative urine output for the current. Our for cleaning or turning the patient. It may be easier to remove the bag and the tubing from the system. First raise the tubing so any urine in the tubing drains into the bag and is added to the current our volume. Then remove the tubing and bags from the sensical urine output system, hang the bag on the bed for adjusting patient position. Make sure the bag stays below the level of the bladder to avoid urinary reflux if necessary. You can move the system with the crossbar handle when ready to resume urine output monitoring hanging the bag and insert the tubing into the tube holder. The system will stabilize and then resume monitoring any urine collected while the patient position was being adjusted will be added to the current hours urine output. There's no need to drain the urine bag before transporting the patient, simply remove the tubing and bag, a message will appear that says the bag has been removed. If your hospital policy requires you to drain the bag prior to patient transport. The empty bag must be rehung on the ring resume button touched and enough time allowed for the system to stabilize with the empty bag. This is necessary to ensure the system can accurately measure accumulated urine output when the patient returns. When the patient returns, move the system back into place, hang the back on the ring and insert the tubing lift tubing to deposit any urine near the system into the back. Then touch the resume button. The sensical urine output system will take a short time to stabilize the sensical urine output system will automatically add any additional urine volume generated by the patient while away The hourly volumes will indicate partial when the bag and tubing had been removed for over 15 minutes and will indicate away if they have been removed for the entire hour. The sensical urine output system software algorithm recognizes weight changes which are true urine output versus false signals or artifact. This results in a short delay in recording the urine once it has entered the bag. However, in some situations such as when a diuretic is administered, there is no pause between bolus of urine output draining into the bag during episodes like this, of continuous large urine output. The sense of urine output system continues to measure urine output but will not automatically update on the screen until drainage settles. However, if you need to record the volume immediately touch the instant update button. The instant update button is normally inactivated and colored green. During episodes of continuous Pulis's of urine output draining into the bag, it automatically becomes active and a notification will be shown touching. The instant update button will override the short, built in delay time and record the volume immediately. The single patient use census. A ring will function for 29 days when the duration is close to 29 days, the system will notify the user if the ring expires, a warning message will appear and have an audible sound and urine output recording will cease until a new ring is connected to the system. To replace the ring, simply removed the used ring and attach a new ring. The system will automatically resume monitoring without losing previous patient data. If there was a time gap between a ring expiring and a new ring attached, the display will indicate zero volume for those hours. The sensical urine output system is indicated to monitor urine output and core bladder temperature when connected to a temperature sensing. Foley catheter Temperature sensing. Foley Catheters are intended for use in conjunction with the sensical urine output system and other compatible 400 series temperature sensing monitors. To provide a measure of core bladder temperature, please consult product label inserts and instructions for use for any indications, contra indications, hazards warnings, cautions and directions for use warnings. The B. D. Sensical urine output system is designed to be used with any barred foley catheter and urological disposables connected to any standard urine drainage bag. 2000 or 2500 mL with or without a urine meter, ensure that the urine drainage bag hanging on the BD sensical ring is always below the level of the patient's bladder to maintain continuous gravity flow of urine into the collection bag. Precaution during system startup and in general practice plugged the beauty sensical urine output system into a wall power supply whenever possible. The system screen will dim when the system is unplugged. To maximize battery life. After using the system on battery backup, plug it back into the wall power supply for recharging and to avoid system shut down due to a drained battery. This information pertains to S. C. C. S 1002. If using SCC S 1001, please refer to the instructions for use for more information for more information on product training and education programs. Contact barred dot e D U at B D dot com Created by